Official recall notice

FDA DeviceClass IINotable

FDA Recall: Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Softw

Recall date
2026-08-12
FDA class
Class II

Why it was recalled

Hazard

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Product description

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Affected products

Lot Code: Model No NCOCO5010; UDI-DI 03760279380179; Lot Numbers 250700050, 251000043, 260300048, 260300049, and 260300050