Official recall notice
FDA Recall: Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Softw
- Recall date
- 2026-08-12
- Brand
- TISSIUM SA
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Product description
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Affected products
Lot Code: Model No NCOCO4010; UDI-DI 03760279380148; Lot Numbers 251000041, 250700048, 260300046, 260400050, and 260300047