FDA DeviceClass IINotable
FDA Recall: Allia Moveo angiographic X-ray system
Description
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Hazard
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Remedy
3 units
Related Recalls
2026-06-17 · FDA_DEVICE
FDA Recall: Allia IGS 5 Pulse angiographic X-ray system
GE Medical Systems, LLC
2026-06-17 · FDA_DEVICE
FDA Recall: Allia IGS 7 Pulse angiographic X-ray system
GE Medical Systems, LLC
2026-06-17 · FDA_DEVICE
FDA Recall: Allia IGS 3 Pulse angiographic X-ray system
GE Medical Systems, LLC
2026-06-17 · FDA_DEVICE
FDA Recall: Allia IGS Pulse angiographic X-ray system
GE Medical Systems, LLC