FDA Recall: Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination of an infant incubat

Date: 2026-06-24
Brand: GE Medical Systems, LLC
Category: Medical Device
FDA Class: Class II

Description

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

Hazard

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

Remedy

23430 units