Official recall notice

FDA DeviceClass IINotable

FDA Recall: GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Produ

Recall date
2026-06-03
FDA class
Class II

Why it was recalled

Hazard

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

Recommended action

What to do

2366 units

Product description

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

Related Recalls