Official recall notice
GE HealthCare Trade Recall (Jun 2026)
Product: GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Recall date
- 2026-06-03
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Product description
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).