Official recall notice

FDA DeviceClass IINotable

Portrait Compute AND Core Services V1.2 Recall (Sep 2026)

Product: PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592

Recall date
2026-09-02
FDA class
Class II

Why it was recalled

Hazard

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

One email a week when there are new recalls. Confirm by email; unsubscribe anytime.

Product description

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

Affected products

GTIN 00198953056752, Serial Numbers: SWT25510001TA, SWT26110010TA, SWT26070001TA, SWT26120008TA.