FDA DeviceClass IINotable
FDA Recall: ICU Medical Plum Solo Precision IV Pump, v1.0.3, Item Number 400010401; infusion pump
Description
In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Hazard
In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Remedy
5059 units
Related Recalls
2026-07-22 · FDA_DEVICE
FDA Recall: ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403;
ICU Medical, Inc.
2026-06-17 · FDA_DEVICE
FDA Recall: Plum Solo Precision IV Pump, 40001-0401
ICU Medical, Inc.
2026-06-17 · FDA_DEVICE
FDA Recall: Plum Duo Precision IV Pump, 40002-0403
ICU Medical, Inc.
2026-06-17 · FDA_DEVICE
FDA Recall: Plum Duo Infusion Pump, 40002-0401
ICU Medical, Inc.