FDA DeviceClass INotable
FDA Recall: Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or int
Description
Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
Hazard
Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
Remedy
168,069 devices globally