Official recall notice

FDA DeviceClass IINotable

Computed Tomography X-ray System Recall (Oct 2026)

Product: Computed Tomography X-ray System, Model: uCT 760 with software

Recall date
2026-10-07
FDA class
Class II

Why it was recalled

Hazard

Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.

One email a week when there are new recalls. Confirm by email; unsubscribe anytime.

Product description

Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.

Affected products

UDI-DI: 06971576831043. Serial Number/Software: 600271/R003.8.4.1810621-Re3300SP1H1_P3

Related Recalls