Official recall notice

FDA DeviceClass IINotable

Intrapulmonary Percussive Ventilation IPV HC Recall (Oct 2026)

Product: Intrapulmonary Percussive Ventilation, IPV HC, Product Number: S00009; Phasitron, P5-HC

Recall date
2026-10-07
FDA class
Class II

Why it was recalled

Hazard

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

One email a week when there are new recalls. Confirm by email; unsubscribe anytime.

Product description

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

Affected products

Product/Catalog Number/UDI-DI: S00009/00849436000358, P5-HC/00849436000525