Official recall notice
FDA Recall: MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ4492
- Recall date
- 2026-06-24
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Recommended action
What to do
342 units
Product description
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.