FDA DeviceClass INotable

FDA Recall: Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if

Date: 2026-06-24
Brand: Datex-Ohmeda Inc
Category: Medical Device
FDA Class: Class I

Description

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

Hazard

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

Remedy

550 (367 US; 183 OUS)

Related Recalls