Official recall notice
Intellijoint Navigation System Recall (Oct 2026)
Product: Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249
- Recall date
- 2026-10-14
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
Product description
A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
Affected products
REF/UDI-DI/MacBook Serial Numbers: 000-0241/00628184002414/C02RP36YG8WP, C02QPCF5G8WN 000-0250/00628184002506/C02RF44TG8WN, C02QW25RG8WL, C02QH355G8WM, C02S22N9G8WP; 000-0249/00628184002490/C02RQ3FUG8WN, C02R4451G8WN