Official recall notice

FDA DeviceClass INotable

FDA Recall: Pediatric care bed; Product Designation: KayserBett IDA;

Recall date
2026-06-10
FDA class
Class I

Why it was recalled

Hazard

If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).

Recommended action

What to do

372 units

Product description

If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).