Official recall notice
FDA Recall: Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04
- Recall date
- 2026-07-08
- Brand
- Medacta Usa Inc
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.
Product description
Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.
Affected products
REF/UDI/Lot: 04.01.10.0970/07630542768910/2585086, 04.01.10.0971/07630542768927/2585087, 04.01.10.0972/07630542768934/2585088, 04.01.10.0973/07630542768941/2585089, 04.01.10.0974/07630542768958/2585090, 04.01.10.0975/07630542768965/2585091, 04.01.10.0976/07630542768972/2585092, 04.01.10.0977/07630542768989/2585093, 04.01.10.0978/07630542768996/2585094, 04.01.10.0979/07630542769009/2585095