Official recall notice

FDA DeviceClass IINotable

FDA Recall: STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quan

Recall date
2026-09-09
FDA class
Class II

Why it was recalled

Hazard

A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified

Product description

A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified

Affected products

UDI : (01)3607450007466(11)250331(17)270331(10)273074(241)00746, Exp 3/31/2027