FDA Recall: Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological
Description
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
Hazard
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
Remedy
68 units