FDA Recall: Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor

Date: 2026-04-15
Brand: Spacelabs Healthcare, Inc.
Category: Medical Device
FDA Class: Class II

Description

Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.

Hazard

Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.

Remedy

1790