Official recall notice

FDA DeviceClass IINotable

FDA Recall: Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor

Recall date
2026-04-15
FDA class
Class II

Why it was recalled

Hazard

Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.

Recommended action

What to do

1790

Product description

Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.

Related Recalls