Official recall notice

FDA DeviceClass IINotable

FDA Recall: MEDLINE Medical Procedure Kits labeled as: NURSING KIT, Medline Kit Number/SKU EDUC1044

Recall date
2026-06-24
FDA class
Class II

Why it was recalled

Hazard

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Recommended action

What to do

730 units

Product description

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Related Recalls