Official recall notice

FDA DeviceClass IINotable

Ingenia 1.5T Recall (Oct 2026) – Philips

Product: Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5T S); Software Versions: R11, R12 and R13;

Recall date
2026-10-07
FDA class
Class II

Why it was recalled

Hazard

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

One email a week when there are new recalls. Confirm by email; unsubscribe anytime.

Product description

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Affected products

1. Model number (REF): 1. 781315 (1.5T); UDI-DI: 00884838099715; Serial Numbers: 61005, 61007, 61014, 61033, 61036, 61047, 61056, 61057, 61076, 61087, 61088, 61105, 61116; 2. Model Number (REF): 781341 (1.5T); UDI-DI: 00884838055322; Serial Numbers: 70042, 70051, 70056, 70062, 70203, 70227, 70229, 70249, 70257, 70262, 70285, 70300, 70311, 70383, 70424, 70432, 70438, 70445, 70453, 70460, 70472, 70588, 70598, 70619, 70638, 70642, 70645, 70647, 70708, 70717, 70724, 70728, 70743, 70750, 70751, 70787, 70801, 70821, 70825, 70836, 70846, 70875, 70876, 70886, 70889, 70896, 70904, 70920, 70957, 70961, 70967, 70970, 70979, 70988, 84049, 84076, 84079, 84080, 84141, 84162, 84170, 84173, 84186, 84205, 84212, 84219, 84233, 84250, 84254, 84266, 84267, 84281, 84289, 84329, 84332, 84335, 84343, 84359, 84360, 84368, 84379, 84388, 84392, 84403, 84415, 84431, 84433, 84436, 84450, 84451, 84467, 84476, 84491, 84500, 84510, 84539, 84549, 84555, 84556, 84568, 84596, 84634, 84635, 84641, 84642, 84652, 84660, 84708, 84713, 84714, 84721, 84741, 84745, 84768, 84804, 84839, 84840, 84853, 84858, 84867, 84869, 84893, 84904, 84908, 84978, 84987, 87068, 87076, 87092, 87196, 87197, 87259; 3. Model Number (REF): 781396 (1.5T); UDI-DI: 00884838009820; Serial Numbers: 41000, 41001, 41021, 41034, 41046, 41056, 41061, 41091, 41116, 41145, 41194, 41231, 41249, 41253, 41254, 41255, 41352, 41354, 41356, 41370, 41400, 41408, 41422, 41428, 41456, 41475, 41492, 41513, 41515, 41548; 4. Model Number (REF): 782115 (1.5T); UDI-DI: 00884838099043; Serial Numbers: 61186, 61191, 61196, 61219, 61228, 61236, 61238, 61272, 61276, 61317, 61382, 61383, 61384, 61386, 61387, 61388, 61390, 61396, 61399, 61401, 61402, 61408, 61411, 61418, 61430; 5. Model Number (REF): 782140 (1.5T); UDI-DI: 00884838108646; Serial Numbers: 45608, 45609; 6. Model Number (REF): 781262 (1.5T CX); UDI-DI: 00884838068445; Serial Numbers: 77047, 77058, 77059, 77120, 77143, 77151, 77158, 77161, 77179, 77180, 77182; 7. Model Number (REF): 781347 (1.5T S); UDI-DI: 00884838068421; Serial Numbers: 81027, 81033, 81145;