FDA DeviceClass IINotable
FDA Recall: Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Nu
Description
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Hazard
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Remedy
444 units
Related Recalls
2026-04-29 · FDA_DEVICE
FDA Recall: Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 45cm. Model Number: TB2-0545FC.
Olympus Corporation of the Americas
2026-04-29 · FDA_DEVICE
FDA Recall: OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.
Olympus Corporation of the Americas
2026-04-29 · FDA_DEVICE
FDA Recall: OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2330.
Olympus Corporation of the Americas
2026-04-29 · FDA_DEVICE
FDA Recall: Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 20cm. Model Number: TB2-0520FC.
Olympus Corporation of the Americas