FDA DeviceClass IIINotable
FDA Recall: Heraeus, PALAMIX duo. Material Number: 66057897.
Description
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Hazard
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Remedy
1,125 units