FDA Recall: Heraeus, PALAMIX duo. Material Number: 66057897.

Recall date
2026-05-06
FDA class
Class III

Why it was recalled

Hazard

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Recommended action

What to do

1,125 units

Product description

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).