Official recall notice

FDA DeviceClass IINotable

Dexcom G7 US Retail Sensor/Transmitter 1-pack Recall (Jul 2026)

Product: Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System

Recall date
2026-07-01
FDA class
Class II

Why it was recalled

Hazard

Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.

One email a week when there are new recalls. Confirm by email; unsubscribe anytime.

Product description

Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.

Affected products

Lot Code: UDI-DI: 00386270004109 Lot Numbers: 1725204004 and 1725069002

Related Recalls