Official recall notice
FDA Recall: Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adapt
- Recall date
- 2026-09-23
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.
Product description
A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.
Affected products
All Serial Numbers: REF: M0067933000 (UDI 00191506002510, UDI (00191506021771); REF: M006793300H0 (UDI 00191506002558); REF: M006793300R0 (UDI 00191506002534), UDI (00191506021788); REF: M006793300D0 (UDI 00191506002527); REF: M006793300Z0 (UDI 00191506002541); UPN M0067933100 (UDI 00191506022549)