Official recall notice

FDA DeviceClass IINotable

FDA Recall: Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adapt

Recall date
2026-09-23
FDA class
Class II

Why it was recalled

Hazard

A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.

Product description

A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.

Affected products

All Serial Numbers: REF: M0067933000 (UDI 00191506002510, UDI (00191506021771); REF: M006793300H0 (UDI 00191506002558); REF: M006793300R0 (UDI 00191506002534), UDI (00191506021788); REF: M006793300D0 (UDI 00191506002527); REF: M006793300Z0 (UDI 00191506002541); UPN M0067933100 (UDI 00191506022549)

Related Recalls