Official recall notice
Fisher Scientific Sure-Vue Urine hCG Recall (Oct 2026)
Product: Fisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine Cassette Test, Model/ Part: FHC-102/9004074, FHC-102/9004073; Alere hCG Cassette (25mIU/mL) in Urine, Model/ Part: FHC-102/92217; Cardinal Health hCG Cassette Rapid test CLIA Waived, Model/ Part: FHC-102/B1077-22; Medline HCG Pregnancy Test Cassette, Model/ Part: FHC-102/MPH12025HCG; Pro Advantage By NDC Urine Pregnancy Cassette Device, Model/ Part: FHC-102/P080063;
- Recall date
- 2026-10-07
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Product description
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Affected products
All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 23900526/00613647000526/1155811105, 2018-01-15; 9004074/10304040024116/1155875704, 2014-06-09; 9004073/10304040024109/1155875704, 2014-06-09; 92217/10815845020109/1156015703, 07/16; B1077-22/20885380020251/1155904704, 9/2014; MPH12025HCG/10888277191877/1156093302, 2015-03-20; P080063/M220P0800633/40001853, 2023-03-30