FDA DeviceClass IINotable

FDA Recall: Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Ava

Date: 2026-06-10
Brand: Bayer Medical Care, Inc.
Category: Medical Device
FDA Class: Class II

Description

Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.

Hazard

Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.

Remedy

11,300 units (1650 US, 9650 OUS)

Related Recalls