Official recall notice
Disposable Tourniquet Cuff Recall (Oct 2026) – ASP Global
Product: Brand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 18" X 4" 2-Line, Red Model/Catalog Number: ASP-564007 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No
- Recall date
- 2026-10-07
- Brand
- ASP Global, LLC.
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Product description
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Affected products
Lot Code: Model Number: ASP-564007 UDI-DI: (01) 00840216106727 All Lots produced from 01SEP2016 to 01MAR2023: P1015STY, P1215STY, O0815STY, O1020STY, N0330STY, N0530STY, N0730STY, M0130STY, M0430STY, M0730STY, M1015STY, M1101STY, M1115STY, L0228STY, L0310STY, L0325STY