FDA DeviceClass IINotable

VERIFY Biological/Integrator Challenge Pack Recall (Oct 2026)

Product: VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007

Recall date
2026-10-14
FDA class
Class II

Why it was recalled

Hazard

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

One email a week when there are new recalls. Confirm by email; unsubscribe anytime.

Product description

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Affected products

UDI: 10724995220331 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027