FDA DeviceClass IINotable

FDA Recall: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Acc

Date: 2026-07-08
Brand: Accuray Incorporated
Category: Medical Device
FDA Class: Class II

Description

In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

Hazard

In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

Remedy

48 units