FDA DeviceClass INotable

FDA Recall: Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Cod

Date: 2026-06-03
Brand: Abiomed, Inc.
Category: Medical Device
FDA Class: Class I

Description

Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.

Hazard

Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.

Remedy

625 units

Related Recalls