Official recall notice
FDA Recall: MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
- Recall date
- 2026-08-26
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Product description
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Affected products
Model/UDI-DI/Serial Numbers: MMT-1711/00763000404550/NG2660048H, NG2663343H, NG3015306H; MMT-1712/00763000404567/NG2628486H, NG2758213H, NG2999651H, NG3349198H, NG3351719H, NG2678842H, NG2831821H, NG2831546H, NG2932027H, NG3049657H, NG3033094H, NG2831538H, NG3738739H