Official recall notice

FDA DeviceClass IINotable

TufTex Over-the-Wire Embolectomy Catheter Recall (Oct 2026)

Product: TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;

Recall date
2026-10-14
FDA class
Class II

Why it was recalled

Hazard

Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.

One email a week when there are new recalls. Confirm by email; unsubscribe anytime.

Product description

Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.

Affected products

Catalog Number: E1651-88; UDI-DI: 00840663100866; Lot Number: XOT00002181;