Official recall notice
FDA Recall: Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B;
- Recall date
- 2026-09-23
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
Blood culture convenience kits contain a component subject to a drug recall that due to an an open or incomplete seal on the packaging of the applicator may compromise the component's sterile barrier and if used may lead to localized skin or soft tissue infection.
Product description
Blood culture convenience kits contain a component subject to a drug recall that due to an an open or incomplete seal on the packaging of the applicator may compromise the component's sterile barrier and if used may lead to localized skin or soft tissue infection.
Affected products
77451/20809160195069/343602(5/31/2027), 349005(10/31/2027), 374030(11/30/2027), 375434(12/31/2027), 377632(4/2/2028), 327194(8/31/2026), 375105(3/31/2028), 336074(3/21/2027); 81834B/20809160403324/326214(7/31/2027); 81834C/20809160466602/347551(8/12/2026), 372502(8/10/2026), 374615(10/21/2026), 376461(10/23/2026), 377088(11/6/2026), 373129(9/3/2026), 374033(9/3/2026)