Official recall notice
FDA DeviceClass IINotable
FDA Recall: ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Mod
- Recall date
- 2026-06-10
- Category
- Medical Device
- FDA class
- Class II
Why it was recalled
Hazard
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
Product description