FDA DeviceClass IINotable

FDA Recall: 3mensio Workstation (Vascular Fenestrated) software

Recall date
2026-07-29
FDA class
Class II

Why it was recalled

Hazard

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

Product description

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

Affected products

Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5