Official recall notice

FDA DeviceClass IINotable

Disposable Tourniquet Cuff Recall (Oct 2026) – ASP Global

Product: Brand Name: Expressaire Product Name:DISPOSABLE TOURNIQUET CUFF, 18" X 4" 1-LINE (CS 18) RED Model/Catalog Number: A70107006 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No

Recall date
2026-10-07
FDA class
Class II

Why it was recalled

Hazard

Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.

One email a week when there are new recalls. Confirm by email; unsubscribe anytime.

Product description

Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.

Affected products

Lot Code: Model Number: A70107006 UDI-DI: (01) 00885914035504 All Lots produced from 01SEP2016 to 01MAR2023: S0823STY, S0922STY, S1214STY, R0601STY, R1101STY, Q0930STY, Q1115STY, P0325STY, P1015STY, O0815STY, N0330STY, N0730STY, M0130STY, M0730STY, M0930STY

Related Recalls