Official recall notice

FDA DeviceClass IIINotable

FDA Recall: Battery Charging Station; Model: 0998-00-0802;

Recall date
2026-03-25
FDA class
Class III

Why it was recalled

Hazard

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.

Recommended action

What to do

28 units (20 US, 8 OUS)

Product description

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.