Official recall notice

FDA DeviceClass IINotable

FDA Recall: Atellica CH Urine Albumin (UAlb). Material Number: 11537225

Recall date
2026-03-11
FDA class
Class II

Why it was recalled

Hazard

Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.

Recommended action

What to do

4,885 units

Product description

Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.