FDA DeviceClass IINotable
FDA Recall: Atellica CH Urine Albumin (UAlb). Material Number: 11537225
Description
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
Hazard
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
Remedy
4,885 units
Related Recalls
2026-05-06 · FDA_DEVICE
FDA Recall: epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN):
Siemens Healthcare Diagnostics Inc
2026-05-06 · FDA_DEVICE
FDA Recall: Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Siemens Healthcare Diagnostics, Inc.
2026-04-08 · FDA_DEVICE
FDA Recall: Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
Siemens Healthcare Diagnostics, Inc.