Official recall notice
FDA DeviceClass INotable
FDA Recall: Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor COR
- Recall date
- 2026-05-13
- Category
- Medical Device
- FDA class
- Class I
Why it was recalled
Hazard
Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.
Recommended action
What to do
7,311 kits
Product description
Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.