FDA DeviceClass IINotable
FDA Recall: HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers.
Description
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Remedy
1,307 units
Related Recalls
2026-04-01 · FDA_DEVICE
FDA Recall: Brand Name: Humeral Head Plate with Angular Stability, with the following description: Proximal Hume
I.T.S. GmbH
2026-04-01 · FDA_DEVICE
FDA Recall: Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers.
I.T.S. GmbH
2026-04-01 · FDA_DEVICE
FDA Recall: I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article num
I.T.S. GmbH
2026-04-01 · FDA_DEVICE
FDA Recall: I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descr
I.T.S. GmbH