FDA DeviceClass IINotable

FDA Recall: MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Date: 2026-03-25
Brand: Howmedica Osteonics Corp.
Category: Medical Device
FDA Class: Class II

Description

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Hazard

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Remedy

1098 units