FDA DeviceClass IINotable

FDA Recall: MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Recall date
2026-03-25
FDA class
Class II

Why it was recalled

Hazard

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Recommended action

What to do

1098 units

Product description

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.