FDA DeviceClass IINotable
FDA Recall: MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Description
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Hazard
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Remedy
1098 units
Related Recalls
2026-05-06 · FDA_DEVICE
FDA Recall: Medline medical procedure kits, containing Medline Neuro Sponges (4077 in total), LAMINECTOMY, CRANI
Medline Industries, LP
2026-05-06 · FDA_DEVICE
FDA Recall: VACUUM BAG SU LARGE, Non-sterile specimen container, Model/Catalog Number: 68409SS
Milestone S.R.L.
2026-05-06 · FDA_DEVICE
FDA Recall: Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. PLASTICS
Medline Industries, LP
2026-05-06 · FDA_DEVICE
FDA Recall: WHILL Model F, with software versions before HMI: 2.23 / MC: 1.05; the device is an indoor/outdoor,
WHILL, INC.