Official recall notice
FDA Recall: Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co
- Recall date
- 2026-04-01
- Category
- Medical Device
- FDA class
- Class I
Why it was recalled
Hazard
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
Recommended action
What to do
28,457
Product description
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.